Dateien in Kürze verfügbar.
Dateien in Kürze verfügbar.
With the filter integrity tester "it-01" and the
filter test box "itb-08" up to 8 small disposable filters or syringe filters can be simultaneously tested in a one test run. All filters are tested in parallel.
The wetting procedure is a part of the integrity testing.
Due to the "itb-08" filter holder design, 50 mm disposable filters as well as LUER LOCK syringe fiters can be inserted.
Installation of the "itb-08" for the filter integrity testing (usually for the Bubble Point Test). The filter test box "itb-08" connects to the pressure vessel on one side and on the other to the outlet
connector of the filter integrity tester "it-01" (see picture below).
After the test starts, the wetting fluid starts to flow through the filters during a selected time. The test result is printed out by the integrated printer in approx. 5 min.
With the filter integrity tester "it-01" and the filter
test box "itb-10" up to 10 small vent filters can be simultaneously tested in a one test run. All filters are tested in parallel.
The Low Pressure Test Box ''itb-LP'' is an accessory for the filter integrity tester ''it-01'' that has been designed for leakage tests on sterile package bags and containers.
It is controlled by the filter integrity tester ''it-01''.
The inlet side (pressure) of the "itb-LP" is linked to the line pressure source (1-3.5 barg).
- The filling valve is disconnected from the filter integrity tester "it-01".
- The cable of the "itb-LP" is plugged into the same connector.
- A PTFE sterile filter capsule is mounted on the outlet side of the
"itb-LP" (Tri-clamp TC 1.5“).
- Sterile bags are fixed to the outlet side of the capsule.
- The test program for the specific leakage test is started on the filter integrity tester "it-01".
- When the test is over, the sterile bags are emptied by the inbuilt vacuum pump.
It is connected and controlled by the "it-01" filter integrity tester.
The wetting procedure of membrane filter is fully automatic as well as the test run.
A suitable optimized wetting procedure is started for every filter test. It minimizes any risks of negative test results, especially with PES membrane filters.
Adapter: Code 7
Connecting plate: tri-clamp TC 4
Material: stainless steel 1.4404
-> Suitable housing bells are on request.
Box "itb-08" for parallel integrity testing of 8 mini filters.
Material: stainless steel 1.4571 e-polished
Vessel Volume: from 5,7L to 19,5L
5.7L - Press. max: 6 bar / Temp. max: 80 °C
10L - Press. max: 6 bar/ Temp. max: 80 °C
19.5L - Press. max: 6 bar/ Temp. max: 80 °C
Connectors:
Inlet Side: Unvalved male Stäubli nipple with G1/4" thread
Outlet Side: Valved female Stäubli connector RBE 03 with G1/4" thread
Material: stainless steel 1.4404
Every pressure vessel is delivered with all necessary documents for the EU Market.
Material: stainless steel 1.4404
Volume measurement range: 0.1 –3.5L
Pressure max.: 8 bar
Temperature max.: 80°C
Connector:
Unvalved Stäubli Male Connector
(yellow coded) (Order-Nr.: St3c-1/8)
Note:
The input value of the reference vessel must be adjusted to the filter integrity tester "it-01".
Material: stainless steel 1.4404
Volume measurement range: 0.3 – 6L
Pressure max.: 8 bar
Temperature max.: 80°C
Connector:
Unvalved Stäubli Male Connector
(yellow coded) (Order-Nr.: St3c-1/8)
Note:
The input value of the reference vessel must be adjusted to the filter integrity tester "it-01".
Sainless steel Reference Vessel it-800AL for Volume Measurement